Key facts
- Brand
- Hanna Instruments
- Model
- HI 3812
- Category
- Test Set
- Documents
- 2
- Languages
- ES · EN
- Last added
- 2023-11-03
Start here

Hanna Instruments HI 3812 Water Hardness Test Kit Instruction Manual
The Hanna Instruments HI 3812 is a water hardness test kit that determines calcium and magnesium hardness in water by EDTA titration, in two ranges: 0 to 30 mg/L (ppm) CaCO3 and 0 to 300 mg/L (ppm) CaCO3. Each kit is supplied with Reactivo 1 (hardness buffer, 30 mL), Reactivo 2 (calmagite indicator, 10 mL), Reactivo 3 (EDTA solution HI3812-0, 120 mL), two beakers (10 mL and 50 mL) and a graduated syringe.
This Spanish-language instruction manual explains how hardness is defined and why it is controlled, the chemistry of the test (buffer to pH 10, calmagite indicator forming a red complex, excess EDTA turning the solution blue at the endpoint), and the step-by-step procedure for the high and low ranges, including the multipliers used to convert the syringe reading into mg/L CaCO3. It closes with specifications, references and health and safety notes. This manual is written in Spanish.
Hanna Instruments HI 3812 specifications
| Range | 0 to 30 mg/L (ppm) CaCO3 and 0 to 300 mg/L (ppm) CaCO3 |
|---|---|
| Minimum increment | 0.3 mg/L in the 0-30 mg/L range; 3 mg/L in the 0-300 mg/L range |
| Analysis method | EDTA titration |
| Sample size | 5 mL and 50 mL (average) |
| Number of analyses | 100 (average) |
| Case dimensions | 200 x 120 x 60 mm |
| Weight | 460 g |
Source: Hanna Instruments HI 3812 Water Hardness Test Kit Instruction Manual
All Hanna Instruments HI 3812 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is included in the HI 3812 test kit?
The kit contains Reactivo 1 (hardness buffer, 30 mL), Reactivo 2 (calmagite indicator, 10 mL), Reactivo 3 (EDTA solution HI3812-0, 120 mL), two beakers (10 mL and 50 mL) and one graduated syringe.
From: Hanna Instruments HI 3812 Water Hardness Test Kit Instruction Manual
How is the water hardness measured?
The hardness level in mg/L (ppm) of calcium carbonate is determined by an EDTA titration.
From: Hanna Instruments HI 3812 Water Hardness Test Kit Instruction Manual
What is the measurement range for water hardness using this kit?
The kit measures a high range of 0 to 300 mg/L (ppm) CaCO3 and a low range of 0.0 to 30.0 mg/L CaCO3.
From: Hanna Instruments HI 3812 Hardness Test Kit Instruction Manual
What is the smallest increment for measurements in the low range?
The smallest increment is 0.3 mg/L when testing within the 0-30 mg/L range.
From: Hanna Instruments HI 3812 Hardness Test Kit Instruction Manual