Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual
The HI 9147 is a waterproof portable dissolved oxygen meter with manual calibration, automatic temperature compensation, salinity compensation and a fixed galvanic probe. Dissolved oxygen readings are shown in mg/L (ppm) or % saturation, and temperature is measured over -5 to 50°C.
The manual explains preliminary checks, controls, specifications, supplied accessories, probe preparation, one-point calibration, measurement, salinity and altitude compensation, probe/membrane maintenance and battery replacement. Two versions are listed: HI 9147-04 with the HI 76409/4 probe (4 m cable) and HI 9147-10 with the HI 76409/10 probe (10 m cable). This manual is written in Portuguese.
Hanna Instruments HI 9147 specifications
| Dissolved oxygen range | 0.0 to 50.0 mg/L (ppm) O2 |
|---|---|
| Oxygen saturation range | 0 to 600% O2 |
| Temperature range | -5.0 to 50.0°C |
| Resolution | 0.1 mg/L or 1% O2; 0.1°C (1°F) |
| Accuracy | ±1% of reading (O2); ±0.2°C (1°F), excluding probe error |
| Calibration | Manual, in saturated air |
| Temperature compensation | Automatic, 0 to 50°C |
| Salinity compensation | 0 to 51 g/L (resolution 1 g/L) |
| Probe | Fixed HI 76409/4 with 4 m cable or HI 76409/10 with 10 m cable |
| Battery | 3 x 1.5V AAA |
| Battery life | 1000 hours of use (backlight off) |
| Auto-off | After approximately 8 minutes |
| Dimensions | 185 x 72 x 36 mm |
| Weight | 450 g with HI 76409/4 probe |
Source: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual
All Hanna Instruments HI 9147 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
- Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual 2 p
- Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual 2 p
- Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual 2 p
- Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual 2 p
Questions answered in these manuals
How is the HI 9147 calibrated?
Calibration is manual in saturated air and is a simple one-point procedure.
From: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual
How often should the probe membrane and electrolyte be replaced?
The manual recommends replacing the membrane every 2 months and the electrolyte once a month.
From: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual
How do I calibrate the HI9147?
Calibration is a simple 1-point procedure performed in air. Switch the meter on, select % mode, and adjust the % trimmer to display 100%.
From: Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual
What is the temperature range for the HI9147?
The temperature range is -5.0 to 50.0 °C (23 to 122 °F).
From: Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual
How long does it take to fully charge the battery?
It takes approximately 14 hours to completely charge the battery.
From: Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual
What is the recommended frequency for replacing the membrane and electrolyte?
For top performance, it is recommended to replace the membrane every 2 months and the electrolyte once a month.
From: Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual