Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual

Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual

User Manual 2 pages · 155 KB · PT

The HI 9147 is a waterproof portable dissolved oxygen meter with manual calibration, automatic temperature compensation, salinity compensation and a fixed galvanic probe. Dissolved oxygen readings are shown in mg/L (ppm) or % saturation, and temperature is measured over -5 to 50°C.

The manual explains preliminary checks, controls, specifications, supplied accessories, probe preparation, one-point calibration, measurement, salinity and altitude compensation, probe/membrane maintenance and battery replacement. Two versions are listed: HI 9147-04 with the HI 76409/4 probe (4 m cable) and HI 9147-10 with the HI 76409/10 probe (10 m cable). This manual is written in Portuguese.

Read online

Hanna Instruments HI 9147 specifications

Dissolved oxygen range0.0 to 50.0 mg/L (ppm) O2
Oxygen saturation range0 to 600% O2
Temperature range-5.0 to 50.0°C
Resolution0.1 mg/L or 1% O2; 0.1°C (1°F)
Accuracy±1% of reading (O2); ±0.2°C (1°F), excluding probe error
CalibrationManual, in saturated air
Temperature compensationAutomatic, 0 to 50°C
Salinity compensation0 to 51 g/L (resolution 1 g/L)
ProbeFixed HI 76409/4 with 4 m cable or HI 76409/10 with 10 m cable
Battery3 x 1.5V AAA
Battery life1000 hours of use (backlight off)
Auto-offAfter approximately 8 minutes
Dimensions185 x 72 x 36 mm
Weight450 g with HI 76409/4 probe

Source: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual

All Hanna Instruments HI 9147 documents by type

Questions answered in these manuals

How is the HI 9147 calibrated?

Calibration is manual in saturated air and is a simple one-point procedure.

From: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual

How often should the probe membrane and electrolyte be replaced?

The manual recommends replacing the membrane every 2 months and the electrolyte once a month.

From: Hanna Instruments HI 9147 Oxycheck Portable Dissolved Oxygen Meter Instruction Manual

How do I calibrate the HI9147?

Calibration is a simple 1-point procedure performed in air. Switch the meter on, select % mode, and adjust the % trimmer to display 100%.

From: Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual

What is the temperature range for the HI9147?

The temperature range is -5.0 to 50.0 °C (23 to 122 °F).

From: Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual

How long does it take to fully charge the battery?

It takes approximately 14 hours to completely charge the battery.

From: Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual

What is the recommended frequency for replacing the membrane and electrolyte?

For top performance, it is recommended to replace the membrane every 2 months and the electrolyte once a month.

From: Hanna Instruments HI 9147 Dissolved Oxygen Meter Instruction Manual