About this manual

This document is Part 1 of a preliminary user manual for the JEOL JXA-733 electron microprobe, prepared by Paul Carpenter of the Division Analytical Facility, Department of Geology, California Institute of Technology. It provides annotated illustrations of the console modules and describes the function of each component.

The text covers the Accelerating Voltage Unit, the Absorbed Electron Microammeter unit, and the Backscattered Electron Image units, including controls such as the GUN FILAMENT knob, GUN BIAS cam switch, LOAD/FILAMENT switch and meter, and OVERLOAD lamp. It also notes operating points related to filament saturation and imaging modes.

What's inside

  • Accelerating Voltage Unit
  • ACCELERATING VOLTAGE SELECT push buttons
  • FINE KV cam switch
  • GUN FILAMENT knob
  • GUN BIAS cam switch
  • LOAD/FILAMENT switch and meter
  • OVERLOAD lamp
  • Absorbed Electron Microammeter (AEM) unit
  • Backscattered Electron Image (BEI and BEI-H) units
  • COMPO and TOPO imaging modes

Specifications

Accelerating voltage range0 to 50 kV
Accelerating voltage selection step10 kV steps
Fine accelerating voltage step1 kV steps
Gun bias steps10 discrete steps
Overload reference current300 uA
AEM range10-5 to 10-12 amps over 15 steps
Vacuum cutoff threshold10-5 torr

Frequently asked questions

What is the accelerating voltage range of the JEOL JXA-733?

The manual states that the instrument can achieve an accelerating voltage range from 0 to 50 KV.

How is the total accelerating voltage set on the console?

It is the sum of the selected ACCELERATING VOLTAGE SELECT push button and the FINE KV cam switch setting.

What happens when the load current exceeds 300 uA?

The accelerating voltage is automatically shut off; the OVERLOAD lamp flashes above 300 uA.

When are the COMPO imaging modes of the BEI units obtained?

When the bse signals from both halves of the bse detector are summed together.