This is part of the electron microscope manual, introducing the part of the electron microscope voltage control
This document is Part 1 of a preliminary user manual for the JEOL JXA-733 electron microprobe, prepared by Paul Carpenter of the Division Analytical Facility, Department of Geology, California Institute of Technology. It provides annotated illustrations of the console modules and describes the function of each component.
The text covers the Accelerating Voltage Unit, the Absorbed Electron Microammeter unit, and the Backscattered Electron Image units, including controls such as the GUN FILAMENT knob, GUN BIAS cam switch, LOAD/FILAMENT switch and meter, and OVERLOAD lamp. It also notes operating points related to filament saturation and imaging modes.
| Accelerating voltage range | 0 to 50 kV |
|---|---|
| Accelerating voltage selection step | 10 kV steps |
| Fine accelerating voltage step | 1 kV steps |
| Gun bias steps | 10 discrete steps |
| Overload reference current | 300 uA |
| AEM range | 10-5 to 10-12 amps over 15 steps |
| Vacuum cutoff threshold | 10-5 torr |
The manual states that the instrument can achieve an accelerating voltage range from 0 to 50 KV.
It is the sum of the selected ACCELERATING VOLTAGE SELECT push button and the FINE KV cam switch setting.
The accelerating voltage is automatically shut off; the OVERLOAD lamp flashes above 300 uA.
When the bse signals from both halves of the bse detector are summed together.