JEOL 2100 Transmission Electron Microscope Quick Check List

This manual provides instructions for using the JEOL 2100 TEM


JEOL 2100 Transmission Electron Microscope Quick Check List - cover page
Brand
JEOL
Category
microscope
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
355 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
B7E0619752104E2DDF7C32BA33D81EBB

About this manual

This document provides a concise quick check list for the JEOL 2100 Transmission Electron Microscope. It outlines the essential steps required to prepare the instrument for operation, including filling the liquid nitrogen reservoir and raising the high tension voltage.

The guide details the procedure for inserting the specimen holder, generating the electron beam, and performing critical alignments such as condenser astigmatism correction, eucentric height adjustment, and objective lens astigmatism correction.

What's inside

  • Quick check list
  • Reservoir filling
  • High tension adjustment
  • Specimen holder insertion
  • Electron beam generation
  • Condenser aperture insertion
  • Eucentric height adjustment
  • Gun shift correction
  • Condenser lens deflection
  • Objective lens astigmatism

Frequently asked questions

What fluid should be used to fill the reservoir?

The reservoir should be filled with LN2 if needed.

To what voltage should the high tension be raised?

The high tension must be raised to 200kV.

How should the specimen holder be inserted into the TEM?

Insert the holder in the airlock, set the switch to pump, wait for the green light, then rotate fully clockwise in two stages and push it in completely.

Download PDF (355 KB)