JEOL 2010F Transmission Electron Microscope Operation Manual

This is a user manual for JEOL 2010F


JEOL 2010F Transmission Electron Microscope Operation Manual - cover page
Brand
JEOL
Category
Industrial Robot
Document type
User Manual
Language
English
Pages
6
File format
PDF
File size
105 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
703F006C5667383D73E2D52C1652B660

About this manual

This manual provides a quick check list and operational procedures for the JEOL 2010F Transmission Electron Microscope. Key initial steps include verifying the accelerating voltage, checking vacuum sequences on the SIP unit, and ensuring the anti-contamination device (ACD) reservoir is filled with liquid nitrogen if needed. Users must ensure the viewing screen cover is on before filling the trap.

What's inside

  • Quick check list
  • Inserting the specimen holder
  • Electron Beam Generation
  • Aligning the TEM
  • Condenser lens adjustment
  • Adjusting Eucentric Height
  • Gun Shift and Condenser Shift Correction
  • Adjusting CL Deflection Coil
  • Centering the Voltage Axis
  • Objective Lens Astigmatism Correction

Frequently asked questions

What must the accelerating voltage be set to?

The accelerating voltage must be at 200 kV.

What is the required vacuum level in the gun?

The vacuum in the gun must be better than x 10-6 Pa.

What should the vacuum level be in the SIP unit?

The vacuum measured on SIP must be better than 2 x 10-5 Pa.

What magnification should be used to center the CL aperture?

Select MAG1 and a suitable magnification, e.g. 40k.