This is a user manual for JEOL 2010F
This manual provides a quick check list and operational procedures for the JEOL 2010F Transmission Electron Microscope. Key initial steps include verifying the accelerating voltage, checking vacuum sequences on the SIP unit, and ensuring the anti-contamination device (ACD) reservoir is filled with liquid nitrogen if needed. Users must ensure the viewing screen cover is on before filling the trap.
The accelerating voltage must be at 200 kV.
The vacuum in the gun must be better than x 10-6 Pa.
The vacuum measured on SIP must be better than 2 x 10-5 Pa.
Select MAG1 and a suitable magnification, e.g. 40k.