This is a manual for using the JEOL-Delta Ver.5 NMR measurement system.
This manual describes how to use the JEOL Delta Ver.5 NMR measurement program, from logging into the workstation and starting Delta to closing the program at the end of a session. It covers opening the Spectrometer Control window, connecting to a spectrometer, setting up the sample and defining it in the software.
The second part explains manual sample control, creating a job, running 1D proton and 1D carbon measurements with example parameter settings, and processing 1D data in the processor window (phasing, referencing, peak detection, integration and printing). A final section lists the steps for shutting Delta down.
| Sample tube diameter | 5 mm |
|---|---|
| Standard sample height | about 4 cm |
| Sample temperature | r.t. to 30 °C |
| 1D Proton spectral range (x_sweep) | 15 ppm |
| 1D Carbon spectral range (x_sweep) | 250 ppm |
Turn on the monitor, press Ctrl+Alt+Del and then enter your username and password.
Open the Spectrometer Control window from the Delta Console, select a spectrometer from the list and click the Connect button.
With a 5 mm diameter sample tube, make the standard sample height about 4 cm.
Delete all jobs, eject the sample tube, delete the sample list, click the Unlink button and then close Delta.