How to linearize arrayed NMR data with the Data Slate on the JEOL Delta Tips 160831.
This document explains how to linearize arrayed data using the Data Slate feature on the JEOL NMR Spectrometer Delta Tips 160831. It describes how to arrange 1D spectra side by side, such as when measuring pulse width or diffusion coefficient. The example shows how to measure pulse width by selecting Expansion – Linearize. The plot of signal intensity vs. pulse width consists of 37 one-dimensional 13C spectra arranged side by side. It also notes that obs_width_ 90 was varied from 1 µs to 19 µs (step 0.5 µs), and that all 1D spectra were stored as a single pseudo-2D data file.
Select Expansion – Linearize to arrange all the 1D spectra side by side.
The plot consists of 37 one-dimensional 13C spectra arranged side by side.
The obs_width_ 90 was varied from 1 µs to 19 µs with a step of 0.5 µs.