JEOL NMR Spectrometer Delta Tips 160831 User Manual

How to linearize arrayed NMR data with the Data Slate on the JEOL Delta Tips 160831.


JEOL NMR Spectrometer Delta Tips 160831 User Manual - cover page
Brand
JEOL
Category
Analyzer
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
895 KB
Published
06 November, 2026
Updated
06 November, 2026
MD5 checksum
1E3FFDF2082F92B2B8B3FA567BD29A42

About this manual

This document explains how to linearize arrayed data using the Data Slate feature on the JEOL NMR Spectrometer Delta Tips 160831. It describes how to arrange 1D spectra side by side, such as when measuring pulse width or diffusion coefficient. The example shows how to measure pulse width by selecting Expansion – Linearize. The plot of signal intensity vs. pulse width consists of 37 one-dimensional 13C spectra arranged side by side. It also notes that obs_width_ 90 was varied from 1 µs to 19 µs (step 0.5 µs), and that all 1D spectra were stored as a single pseudo-2D data file.

What's inside

  • How to linearize arrayed data
  • Data Slate
  • Visualization of Arrayed Data

Frequently asked questions

How do you linearize arrayed data in Data Slate?

Select Expansion – Linearize to arrange all the 1D spectra side by side.

How is the plot of signal intensity vs. pulse width arranged?

The plot consists of 37 one-dimensional 13C spectra arranged side by side.

What was varied in the pulse width example?

The obs_width_ 90 was varied from 1 µs to 19 µs with a step of 0.5 µs.