JEOL JEM-2010 TEM Basic Operation Handbook

This manual introduces the basic operating methods of JEOL JEM-2010 TEM, including sample insertion, beam positioning, basic operation, replacing specimens, detailed operation, detailed alignment, Gatan GIF startup and alignment.


JEOL JEM-2010 TEM Basic Operation Handbook - cover page
Brand
JEOL
Category
microscope
Document type
User Manual
Language
English
Pages
27
File format
PDF
File size
616 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
07D7BD5B468B1F655993748A0153D938

About this manual

The JEOL JEM-2010 Handbook - Basic Operation (Second Edition) is a 27-page laboratory guide from the NTNU Department of Materials Technology Electron Microscopy Laboratories. It is written for trained users and stresses cleanliness, responsible use, and immediate reporting of instrument problems.

The contents cover guidelines, sample insertion and filament procedure, finding the beam, basic operation in DIFF, BF, DF, photo, CBD, NBD and EDS modes, changing the specimen or ending a session, more detailed instruction and alignment, and Gatan GIF start-up and alignment.

What's inside

  • Guidelines
  • Sample Insertion and Filament Procedure
  • How to find beam (short form)
  • Basic operating: DIFF, BF, DF, Photo, CBD, NBD and EDS
  • Changing specimen or ending session
  • More detailed instruction
  • More detailed alignment
  • Gatan GIF start-up and alignment

Specifications

Starting accelerating voltage160 kV
Final accelerating voltage200 kV
Filament typeLaB6 single crystal
Filament run time10 minutes for first user of the day; 6 minutes if already used the same day; 3 minutes when changing specimen
Double tilt holder materialBeryllium (Be)

Frequently asked questions

What should I check before inserting the sample holder into the TEM?

Check that the sample position is x = 0, y = 0, sample height z = 0, and tilt is 0 degrees on the TEM PC screen.

How long should the filament be run?

Run for 10 minutes if you are the first user of the day, 6 minutes if the TEM has already been used that day, and 3 minutes when changing specimen.

What safety equipment is required when filling the cold finger trap with LN2?

Use gloves, eye protection, and a funnel.

How do I know the specimen holder has been inserted far enough?

Insert it until it stops and two audible clicks are heard.