This is an instruction manual on how to use JEOL JEM-1400 microscope
This document is a Microscope Usage protocol for the JEOL JEM-1400 at the Robert P. Apkarian Integrated EM Core (IEMC). It is intended as instruction guidelines for users who have completed training and been approved by IEMC staff, and it assumes knowledge of the microscope controls.
The protocol walks through pre-use checks, inserting the sample, gun alignment, eucentric height, objective lens aperture centering and stigmation, sample removal, end-of-session shutdown, voltage centering and Gatan CCD gain reference.
| Accelerating voltage | 80 kV usually selected; 100 kV to 120 kV may be optimal for some samples |
|---|---|
| Dark current | 44 µA +/- 2 µA at 80 kV; 66 µA +/- 2 µA at 120 kV |
| Vacuum | about 13 in the units shown |
| End-of-day voltage and magnification | 120 kV and 30,000 x (30kx) |
An accelerating voltage of 80 kV is usually selected; 100 kV to 120 kV may be optimal for some samples.
Select Control (C), then Stage, then Stage Neutral; the stage coordinates should be close to 0.0.
It should be left at 120 kV and a magnification of 30,000 x (30kx).
It should stay at 44 µA +/- 2 µA.