Use JEOL AccuTOF™-DART™ detection system
This training guide describes the JEOL AccuTOF-DART system and its Direct Analysis in Real Time (DART) ion source. It explains the DART ion source schematic, ionization mechanisms for positive and negative ions, and the main parts of the AccuTOF spectrometer, including ion transport, analyzer and detection system.
The guide also gives procedures for system start-up and shutdown, conditioning the MCP, tuning the AccuTOF with reserpine, acquiring and processing data, configuring negative ion analysis, determining accurate mass, and troubleshooting common hardware and MassCenter issues.
| Carrier gas | Usually helium |
|---|---|
| He(23S) excitation energy | 19.8 eV |
| Analyzer | Orthogonal, two-stage acceleration, time-of-flight mass spectrometer incorporating a single-stage reflectron |
| Detection system | Two micro channel plates (MCP) with a continuous digital averager (ADC) |
| Ion transport | Quadrupole ion guide with three focusing lenses |
| Standard ion sources | DART and orthogonal spray ESI ion sources |
It usually uses helium; the gas is introduced and passed by a needle electrode.
It filters out charged particles so that only metastable species continue through the source.
It uses an orthogonal, two-stage acceleration time-of-flight mass spectrometer with a single-stage reflectron.
It shows the error message “MassCenter is unstable. Please restart and try again.”