JEOL AccuTOF-DART Mass Spectrometer Training Guide

Use JEOL AccuTOF™-DART™ detection system


JEOL AccuTOF-DART Mass Spectrometer Training Guide - cover page
Brand
JEOL
Category
Analyzers
Document type
User Manual
Language
English
Pages
65
File format
PDF
File size
32.8 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
AA4A3745FE3BC959C4B178436BF4D775

About this manual

This training guide describes the JEOL AccuTOF-DART system and its Direct Analysis in Real Time (DART) ion source. It explains the DART ion source schematic, ionization mechanisms for positive and negative ions, and the main parts of the AccuTOF spectrometer, including ion transport, analyzer and detection system.

The guide also gives procedures for system start-up and shutdown, conditioning the MCP, tuning the AccuTOF with reserpine, acquiring and processing data, configuring negative ion analysis, determining accurate mass, and troubleshooting common hardware and MassCenter issues.

What's inside

  • Basic Hardware Overview
  • DART Ion Source
  • DART Ionization Mechanisms
  • AccuTOF Spectrometer
  • System Start-up / Shut-down
  • Conditioning the MCP
  • Tuning the AccuTOF
  • Data Acquisition and Processing
  • Configuring DART for Negative Ion Analysis
  • Troubleshooting

Specifications

Carrier gasUsually helium
He(23S) excitation energy19.8 eV
AnalyzerOrthogonal, two-stage acceleration, time-of-flight mass spectrometer incorporating a single-stage reflectron
Detection systemTwo micro channel plates (MCP) with a continuous digital averager (ADC)
Ion transportQuadrupole ion guide with three focusing lenses
Standard ion sourcesDART and orthogonal spray ESI ion sources

Frequently asked questions

What gas does the DART ion source usually use?

It usually uses helium; the gas is introduced and passed by a needle electrode.

What is the purpose of electrode 1 in the DART source?

It filters out charged particles so that only metastable species continue through the source.

What type of analyzer does the AccuTOF use?

It uses an orthogonal, two-stage acceleration time-of-flight mass spectrometer with a single-stage reflectron.

What does MassCenter show if it is unstable?

It shows the error message “MassCenter is unstable. Please restart and try again.”