This document introduces the accuracy and precision technical overview of the Agilent Bravo Automated Liquid Handling Platform's 384-channel 30 µL tip
This document provides a technical overview of the Agilent Bravo Automated Liquid Handling Platform, specifically detailing the accuracy and precision performance when used with a 384-Channel ST Disposable Tip Head and 30 uL tips. The platform is designed for automated liquid handling in laboratories, capable of accommodating 96-, 384-, or 1536-well SBS standard microplates across nine deck positions.
| Tip Volume Tested | 30 uL |
|---|---|
| Precision at 0.5 pL | 2.7% CV |
| Accuracy at 0.5 pL | +5.0% |
| Precision at 1.0 pL | 2.0% CV |
| Accuracy at 1.0 pL | +1.0% |
| Precision at 5.0 pL | 1.1% CV |
| Accuracy at 5.0 pL | +1.0% |
The precision is 2.7% CV.
The accuracy is +1.0%.
It is compatible with 96LT 200 pL, 96LT 250 pL, and 384ST 10 uL, 30 uL, and 70 pL plates.