Agilent SampliQ OPT Solid-Phase Extraction Cartridges Technical Note

Agilent SampliQ OPT uses Optimized Polymer Technology (OPT), a novel polyamide chemistry (patent pending). The result is a resin that exhibits retention for both polar and nonpolar compounds based on the balanced hydrophilic and hydrophobic properties of the polymer.


Agilent SampliQ OPT Solid-Phase Extraction Cartridges Technical Note - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
163 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
5FD0A1A296E8B14E3A2D961B41439275

About this manual

Agilent introduces the SampliQ OPT solid-phase extraction (SPE) cartridges, which utilize a novel polyamide chemistry known as Optimized Polymer Technology (OPT). This technology provides a balanced hydrophilic/lipophilic character, allowing for the retention of both polar and nonpolar compounds, including basic, neutral, and acidic substances. The document highlights that these cartridges are robust, stable in pH ranges from 0 to 14, and capable of withstanding drying between conditioning and equilibration steps without affecting reproducibility.

What's inside

  • General Description
  • Quality Controls
  • Operational Guidelines
  • Conditioning Step
  • Loading Step
  • Washing Step
  • Elution Step
  • Performance
  • Part Number Description

Specifications

pH Stability Range0 to 14
Cartridge Volume (Part 5982-3013)1 mL
Cartridge Volume (Part 5982-3036)3 mL
Cartridge Volume (Part 5982-3067)6 mL
Recommended Flow Rate1 mL/min

Frequently asked questions

Can the SampliQ OPT cartridge run dry between conditioning and equilibration?

Yes, the cartridges are robust and can run dry between the methanol conditioning and water equilibration steps without affecting reproducibility.

What is the recommended flow rate for loading samples?

The flow through the cartridge should be no faster than 1 mL/minute for the loading step.

Which compounds are not effectively retained on SampliQ OPT?

Compounds that are both hydrophilic (low log P) and strongly basic (pKa > 10) or strongly acidic (pKa < 3) will not be effectively retained.