Agilent Technologies 1200 Series Variable Wavelength Detectors Performance Characteristics Technical Note

Agilent 1200 Series Variable Wavelength Detectors Performance Characteristics: Faster results, improved sensitivity and absolute data security


Agilent Technologies 1200 Series Variable Wavelength Detectors Performance Characteristics Technical Note - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
581 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
926E7183958FA8692B3A39CCF77697CE

About this manual

This document presents a technical evaluation of the Agilent 1200 Series Variable Wavelength Detectors (VWD) and the high-speed VWD SL Plus. It compares their performance against an earlier model, highlighting improvements in optical design, data sampling rates, and baseline stability for applications ranging from conventional to ultra-fast HPLC.

Key performance metrics analyzed include noise levels, drift behavior, limit of detection (LOD) using anthracene, signal-to-noise ratios, and the ability to detect impurities at low concentrations. The text details how the new series achieves superior results through features like built-in electronic temperature control and high sampling rates up to 160 Hz.

What's inside

  • Introduction
  • Optical design
  • Data sampling rates
  • Noise and drift
  • Limit of detection of anthracene
  • Detection of impurities
  • Performance comparison between conventional and ultra-fast LC

Specifications

Wavelength range (Deuterium lamp)190 to 600 nm
Maximum sampling rate (VWD SL Plus)up to 160 Hz
Sampling rate (Standard VWD)up to 20 Hz
Noise improvement factor vs earlier model3.4 times better
Drift improvement factor vs earlier model2.4-fold improvement

Frequently asked questions

What is the limit of detection for anthracene using the Agilent 1200 Series VWD?

The limit of detection was determined to be 0.228 pg based on a 5 pg injection.

Which detector is recommended for ultra-fast LC applications?

It is recommended to use the Agilent 1200 Series VWD SL Plus for rapid resolution and ultra-fast LC.

What features ensure data security in the Agilent 1200 Series VWD SL Plus?

The Data Recovery Card (DRC) provides unique "data-never-lost insurance".

Download PDF (581 KB)

More Agilent Manuals


View all Agilent manuals →