Agilent Velocity11 Bravo Liquid Handling Platform Technical Note

This document describes the features and performance of the Velocity11 Bravo liquid handling platform, as well as the advantages of using this platform for automated liquid handling.


Agilent Velocity11 Bravo Liquid Handling Platform Technical Note - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
157 KB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
7363C99B6697F506F4E3798000DF0945

About this manual

This document serves as a technical note for the Agilent Velocity11 Bravo liquid handling platform, specifically evaluating the performance of its 96-channel ST disposable tip head with 10 uL tips. The platform is designed for automated liquid handling in laboratories, offering nine deck positions that can accommodate various microplate formats and be configured with heating, cooling, or shaking stations.

The note describes a testing method using a tartrazine/DMSO solution to measure accuracy and precision across a volume range of 0.3 to 2.0 uL. Results indicate that the system achieves a coefficient of variation (CV) of 3.1% and an accuracy of +1.5% at 0.3 uL, with performance meeting or exceeding 2.0 uL specifications at higher volumes. The document concludes that the platform is ideal for applications requiring precise dispensing at low volumes, such as genomic applications and cell-based assays.

What's inside

  • Introduction
  • Bravo Features
  • Accuracy and Precision Testing
  • Materials
  • Method
  • Results
  • Conclusion

Specifications

Tip Volume Range0.3-10 uL
Deck Positionsnine
Accuracy at 0.3 uL+1.5%
Precision at 0.3 uL3.1% CV
Accuracy at 0.5 uL+0.8%
Precision at 0.5 uL2.3% CV
Accuracy at 2.0 uL+0.8%
Precision at 2.0 uL1.1% CV

Frequently asked questions

What is the accuracy and precision of the Bravo with 10 uL tips at 0.3 uL?

The accuracy is +1.5% and the precision is 3.1% CV.

Can the Bravo be used inside laminar flow hoods?

Yes, it is designed to fit inside some models of laminar flow hoods and retain laminar flow.

What microplate formats does the Bravo accommodate?

It accommodates any 96-, 384-, or 1536-well SBS standard microplates.