This document introduces the Agilent Bravo Automated Liquid Handling Platform 96-Channel 10 µL Tip Accuracy and Precision Technical Overview.
This document provides a technical overview of the Agilent Bravo Automated Liquid Handling Platform, specifically evaluating its accuracy and precision when used with a 96-Channel ST Disposable Tip Head. The evaluation focuses on low-volume dispensing ranging from 0.3 pL to 2.0 pL, demonstrating performance that meets or exceeds standard laboratory requirements for automated liquid handling.
| Tip Volume Range Tested | 0.3-2 pL (Performance >2 uL meets or exceeds 2 uL performance) |
|---|---|
| Target Accuracy | +10% of desired volume |
| Target Precision (CV) | 5% |
| Deck Positions | 9 |
| Compatible Tip Heads | 8-, 16-, 96-, or 384-channel fixed tip or disposable tip heads |
The product meets or exceeds a Coefficient of Variance (CV) of 5% and an accuracy of +10% of the desired volume across the practical volume range.
The 384ST 10 uL tip is compatible with 96-Well, 1536-Well, and 384-ST microplates.
Yes, it is designed to fit inside some models of laminar flow hoods while retaining laminar flow.