Agilent ZORBAX SB-C18 Rapid-Resolution High-Speed Separation of Analgesics Application Note

The document describes a method for high-speed separation of analgesics using smaller particles and column dimensions along with high temperature to resolve seven analgesics in less than one minute. The use of high temperature reduces run times. ZORBAX SB-C18 can operate at higher temperatures and lower pH than other commercial reversed phase columns.


Agilent ZORBAX SB-C18 Rapid-Resolution High-Speed Separation of Analgesics Application Note - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
402 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
06ED251DB060450E196E7E921D7F7DFE

About this manual

This document presents a high-speed liquid chromatography method designed to separate seven specific analgesics in under one minute. The method utilizes high temperature and a specialized column with small particle sizes to achieve rapid resolution without requiring modifications to the instrument. The approach is particularly suitable for pharmaceutical analysis and can be adapted for LC/MS applications by using a volatile buffer instead of an ion-pair mobile phase. Key benefits include reduced analysis time, maintained resolution, and reduced solvent usage.

What's inside

  • Application
  • Highlights
  • Conditions
  • Mobile Phase
  • Column Specifications

Specifications

Flow Rate2 mL/min
Injection Volume1 µL
Column Temperature70°C
Detector Wavelength275 nm

Frequently asked questions

What is the analysis time for this method?

The method resolves seven analgesics in less than one minute.

Can this method be used for LC/MS?

Yes, the volatile buffer can be used in place of the ion-pair mobile phase for LC/MS.