Agilent ZORBAX SB-C8 Rapid Analysis of Dexamethasone USP Method

This document describes the rapid analysis of dexamethasone using the USP method. By choosing a ZORBAX Rapid-Resolution column configuration, analysis time is reduced by 70% and solvent consumption is subsequently reduced. Improved performance can be achieved by selecting an alternative column configuration.


Agilent ZORBAX SB-C8 Rapid Analysis of Dexamethasone USP Method - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
557 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
39474E1A64528BAA8A2E38B012F015EC

About this manual

The document presents an application note detailing a High-Performance Liquid Chromatography (HPLC) method for the pharmaceutical analysis of Dexamethasone. It highlights the use of Agilent ZORBAX SB-C8 columns to reduce analysis time by 70% and solvent consumption compared to standard methods.

What's inside

  • Application
  • Highlights
  • Column Configuration
  • Mobile Phase
  • Injection Conditions

Frequently asked questions

What is the mobile phase composition for the Dexamethasone analysis?

The method uses an isocratic mobile phase consisting of water and acetonitrile (ACN) with 30% B.

What are the injection volume and flow rate requirements?

The procedure requires an injection of 10 µL and a flow rate of 2 mL/min.

What wavelength is used for UV detection?

Detection is performed using UV at a wavelength of 254 nm.