Agilent 1100 LC/MSD LSD Limit-of-Detection Application Note

This document describes a method for LSD detection using liquid chromatography-mass spectrometry. The method features rapid separation, good peak shape, and excellent sensitivity. The use of an internal standard, LAMPA, enables quantitative analysis of LSD. The method is applicable in the pharmaceutical field.


Agilent 1100 LC/MSD LSD Limit-of-Detection Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
174 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
7A5AACF8E55963DD321594A4CD2DBAB2

About this manual

This document presents a rapid isocratic separation method developed for the detection of LSD and the internal standard LAMPA using an Agilent 1100 LC/MSD system. The method utilizes a ZORBAX Eclipse XDB-C8 column and is optimized for API-ES-MS, demonstrating excellent sensitivity with a limit of detection (LOD) of approximately 4.5 pg on column for LSD. The approach is designed to be legally defensible by including two confirming ions for analysis.

What's inside

  • LSD Limit-of-Detection
  • LC-MS Method
  • Internal Standard LAMPA
  • ZORBAX Eclipse XDB-C8 Column
  • API-ES-MS Conditions
  • SIM Mode Ions
  • Linearity of Detection
  • Application Pharmaceutical

Specifications

Limit of Detection (LSD)4.5 pg on column
Internal StandardLAMPA
ColumnZORBAX Eclipse XDB-C8, 2.1 x 50 mm, 5µm
Mobile Phase15 : 85, ACN : H2O containing 10 mM ammonium formate, pH 3.7
Flow Rate0.3 mL/min
Injection Volume30 µl
Column Temperature30°C
SIM Ions324.2, 223.1, 208.1
Linearity (r2)0.99959

Frequently asked questions

What is the limit of detection for LSD in this method?

The limit of detection is about 4.5 pg on column.

Which column is used for the separation?

The method uses a ZORBAX Eclipse XDB-C8 column with dimensions 2.1 x 50 mm and 5µm particle size.

What ions are monitored in SIM mode?

The method monitors ions at m/z 324.2, 223.1, and 208.1.

What is the correlation coefficient for the linearity of LSD detection?

The linearity of detection for LSD and LAMPA has an r2 value of 0.99959.