Agilent ZORBAX Eclipse XDB-C18 HPLC Application Note for Pramoxine Analysis

The document describes the method and advantages of using ZORBAX Eclipse XDB-C18 for HPLC analysis. By using ZORBAX Eclipse XDB-C18, improved peak shape, higher separation efficiency, greater sensitivity, reduced analysis time, reduced back pressure, and reduced solvent use per analysis can be achieved. The performance of this product surpasses the requirements of USP, with a higher number of theoretical plates and a closer tailing factor to the ideal value.


Agilent ZORBAX Eclipse XDB-C18 HPLC Application Note for Pramoxine Analysis - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
346 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
5789187F0D6AE06FE55553A0B25A6612

About this manual

This document presents an application method for the high-performance liquid chromatography (HPLC) analysis of Pramoxine, a local anesthetic. It compares the performance of the original USP method utilizing older L1 columns against a modern alternative using the Agilent ZORBAX Eclipse XDB-C18 column.

The text highlights that the ZORBAX Eclipse XDB-C18 offers superior efficiency and peak shape compared to traditional 10 µm C18 or L1 columns. Key benefits include greater sensitivity, reduced analysis time, lower back pressure, and shorter solvent usage due to the availability of shorter column lengths (e.g., 150 mm). The document provides specific chromatographic conditions used for this comparison.

What's inside

  • Application
  • HPLC Analysis
  • Pramoxine HCl
  • L1 Column Comparison
  • ZORBAX Eclipse XDB-C18
  • Chromatographic Conditions
  • Peak Shape Improvement
  • Theoretical Plates Efficiency
  • Tailing Factor
  • Agilent Technologies

Specifications

Column TypeZORBAX Eclipse XDB-C18
Particle Size Options3.5 and 5.0 um
Column Dimensions (Modern)4.6 x 150 mm
Flow Rate (Modern Column)1 mL / min.
Theoretical Plates (N)12,200
Mobile Phase Composition55% ACN : 45% 50 mM KH2PO4, pH 7.5
Detection Wavelength224 nm
Column Part Number993967-902

Frequently asked questions

What are the benefits of using ZORBAX Eclipse XDB-C18 for pramoxine analysis?

Using this column offers greater sensitivity, reduced analysis time, reduced back pressure, and shorter length which reduces solvent use per analysis.

How does the efficiency of ZORBAX Eclipse XDB-C18 compare to the original L1 column?

The number of theoretical plates is more than 8 times greater on the modern column (12,200 vs. 1500).