Agilent Analysis of Carbamazepine and Phenytoin in Plasma Application Note

The document describes a method for analyzing carbamazepine and phenytoin in plasma. The analysis is performed using online extraction and a ZORBAX SB-CN column, allowing for fast and direct analysis of these anticonvulsant drugs.


Agilent Analysis of Carbamazepine and Phenytoin in Plasma Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
149 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
80F884408F683E50CEEAF1F9DAF6DEA4

About this manual

This document details an analytical procedure for determining the levels of carbamazepine and phenytoin in plasma samples. The method utilizes on-line sample preparation and concentration via column switching to provide a fast and direct approach for clinical settings where these anticonvulsant drugs are assayed. The final separation is achieved using a ZORBAX SB-CN column, which operates optimally at low pH.

What's inside

  • Application
  • Highlights
  • Conditions
  • Copyright

Specifications

ColumnZORBAX SB-CN, 4.6 x 150 mm, 5um
Mobile Phase28% ACN in 116 mM sodium phosphate buffer, pH 2.8
Flow Rate1.0 ml/min
DetectionUV 210 nm
Sample Volume200 uL

Frequently asked questions

What is the intended use of this procedure?

This procedure is for Investigational / Research only and is not for in vitro diagnostic procedures.

What column is used for the separation?

The analysis uses a ZORBAX SB-CN column.

Download PDF (149 KB)

More Agilent Manuals