Agilent G6410A Triple Quadrupole Mass Spectrometer Amphetamine Analysis Application Note

A fast, sensitive, and validated technique for confirming the presence of amphetamine drugs in whole blood using the Agilent G6410A Triple Quadrupole Mass Spec-trometer (QQQ) is presented. Excellent linearity is demon- strated over the range of approximately 15 to 1,000 ng/mL. The amphetamine drugs analyzed in this work include amphetamine, methamphetamine, methyl-enedioxyamphetamine (MDA), and methylenedioxymeth-amphetamine (MDMA) in blood. The drugs have been prepared using an extractive alkylation technique. The sample preparation is then followed by reverse-phase LC/MS/MS using a 1.8-µm particle size C18 column for high chromatographic resolution with a high-speed sepa-ration. As a result, elution times for both analytes and internal standards are all less than or equal to 3.6 min- utes.


Agilent G6410A Triple Quadrupole Mass Spectrometer Amphetamine Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
210 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
G6410A
MD5 checksum
1A54E9508AB8BF448F258D97796DDE42

About this manual

This document presents a fast, sensitive, and validated technique for confirming the presence of amphetamine drugs in whole blood. The method utilizes an extractive alkylation technique followed by reverse-phase LC/MS/MS using the Agilent G6410A Triple Quadrupole Mass Spectrometer with electrospray ionization.

The analysis covers amphetamine, methamphetamine, MDA, and MDMA in a concentration range of approximately 15 to 1,000 ng/mL. The method demonstrates excellent linearity (R² > 0.999) with elution times for analytes and internal standards less than or equal to 3.6 minutes.

What's inside

  • Introduction
  • Reagents
  • Standards
  • Sample Preparation
  • LC/MS/MS Instrumentation
  • LC Conditions
  • MS Conditions
  • MRM Settings
  • Results and Discussion

Specifications

Analyte Range15 to 1,000 ng/mL
Elution Time<= 3.6 minutes
Column Temperature60°C
Flow Rate0.7 mL/min
Injection Volume2uL
Ionization ModePositive ESI

Frequently asked questions

Which amphetamine drugs are analyzed in this method?

The method analyzes amphetamine, methamphetamine, methylenedioxyamphetamine (MDA), and methylenedioxymethamphetamine (MDMA).

What is the tolerance for ion ratio confirmation?

The overall ion ratio of the qualifier to the quantifier ions must fall within a +/- 20% tolerance.

What software controls the system?

System control and data analysis are provided by Agilent MassHunter B.01.01 software.