Agilent 6410 MS Application Note: Determination of Cocaine and Metabolites in Hair

A quantitative analytical procedure for the determination of cocaine, benzoylecgonine, cocaethylene, and norco- caine in hair has been developed and validated. The hair samples were washed, incubated, and any drugs present were quantified using mixed-mode solid-phase extraction and liquid chromatography with tandem mass spectro- metric detection in positive atmospheric pressure chemi- cal ionization mode. For confirmation, two transitions were monitored and one ion ratio was determined, which was within 20% of that of the known calibration stan- dards. The monitoring of the qualifying transition and requirement for its presence within a specific ratio to the primary ion limited the sensitivity of the assay, particu- larly for benzoylecgonine; however, the additional confi- dence in the final result as well as forensic defensibility were considered to be of greater importance. Even with simultaneous monitoring, the concentrations proposed by the United States federal guidelines for hair analysis were achieved. The limits of quantitation were 50 pg/mg; the limit of detection was 25 pg/mg. The intra-day precision Determination of Cocaine, Benzoylecgo- nine, Cocaethylene, and Norcocaine in Human Hair Using Solid-Phase Extraction and Liquid Chromatograp


Agilent 6410 MS Application Note: Determination of Cocaine and Metabolites in Hair - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
387 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
47BA49DC57B4724F0876E32EB2E57137

About this manual

The document details a quantitative analytical procedure developed by Immunalysis Corporation and Agilent Technologies to determine specific drugs in human hair. The method utilizes mixed-mode solid-phase extraction followed by liquid chromatography with tandem mass spectrometric detection (LC/MS/MS) in positive atmospheric pressure chemical ionization mode.

Key performance metrics achieved include limits of quantitation at 50 pg/mg and a limit of detection of 25 pg/mg. The procedure involves monitoring two transitions for each analyte to ensure forensic defensibility, with ion ratios required within 20% of calibration standards.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Sample Preparation
  • Calibrators
  • Data Analysis
  • Selectivity
  • Linearity and Sensitivity
  • Precision
  • Stability

Specifications

Limit of Quantitation (LOQ)50 pg/mg
Limit of Detection (LOD)25 pg/mg
Column TypeAgilent ZORBAX XDB-C18, 46 x 50 mm, 1.8 ym
Flow Rate0.9 mL/min
Injection Volume2 uL
Column Temperature40°C

Frequently asked questions

What is the limit of quantitation for this method?

The limits of quantitation were achieved at 50 pg/mg.

Which drugs are analyzed in this procedure?

The method analyzes cocaine, benzoylecgonine, cocaethylene, and norcocaine.

What ionization mode is used?

The method uses positive atmospheric pressure chemical ionization (APCI) mode.