Agilent 1200 Series RRLC 6410 LC Triple Quad Benzodiazepine Analysis Application Note

A quantitative analytical procedure for the determination of benzodiazepines and metabolites in hair has been developed and validated. The hair samples were washed, incubated, and any drugs present were quantified using mixed mode solid-phase extraction and liquid chromatography with tandem mass spectrometric detection (LC/MS/MS) in positive electrospray ionization mode.


Agilent 1200 Series RRLC 6410 LC Triple Quad Benzodiazepine Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
457 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
6091D49BC3B1EEE1B1460A8BD4B4D3E2

About this manual

This document details a quantitative analytical procedure for determining benzodiazepines and their metabolites in hair. The method utilizes mixed-mode solid-phase extraction followed by liquid chromatography with tandem mass spectrometric detection (LC/MS/MS) in positive electrospray ionization mode.

The procedure covers the analysis of 14 benzodiazepines and 5 metabolites, with a limit of quantitation of 50 pg/mg. The method was validated for linearity, precision, and accuracy over a concentration range of 50 to 1,000 pg/mg. Key features include the simultaneous monitoring of qualifying ions to ensure result confirmation, a critical requirement for forensic analysis.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • Extraction Procedure
  • Analytical Procedure
  • LC Conditions
  • MS Conditions
  • MRM Transitions
  • Method Validation

Specifications

Limit of Quantitation50 pg/mg of hair
Analysis Range50 to 1,000 pg/mg
InstrumentAgilent 1200 Series RRLC; 6410 LC Triple Quadrupole Mass Spectrometer
ColumnZORBAX RRHT Eclipse XDB C18, 4.6 mm x 50 mm x 1.8 um
Ionization ModeElectrospray ESI positive mode

Frequently asked questions

What is the limit of quantitation for this benzodiazepine hair analysis method?

The limit of quantitation is 50 pg/mg of hair.

Which Agilent instruments are used for this analysis?

The analysis uses an Agilent 1200 Series RRLC coupled with a 6410 LC Triple Quadrupole Mass Spectrometer.

How many benzodiazepines and metabolites are analyzed simultaneously?

The procedure detects 14 benzodiazepines and 5 metabolites in hair.