Agilent 6410 Determination of Benzodiazepines in Urine and Blood Using LC/MS/MS

A rapid, simple, highly sensitive procedure for the simultaneous analysis of 14 benzodiazepines and six metabolites in urine and blood, using the Agilent 6410 Triple Quadrupole Mass Spectrometer in electrospray mode, is described.


Agilent 6410 Determination of Benzodiazepines in Urine and Blood Using LC/MS/MS - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
535 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
D51BCDF57A6CFA88E3C1116CD308B447

About this manual

The document outlines a forensic procedure for the simultaneous analysis of benzodiazepines and their metabolites in urine and blood samples. The method utilizes the Agilent 6410 Triple Quadrupole Mass Spectrometer in electrospray mode, featuring unique software capabilities to monitor qualifying ions and calculate ion ratios for result confirmation.

Detailed sample preparation protocols are provided for both urine and blood, including specific extraction and cleanup steps. The analytical procedure specifies the use of an Agilent 1200 Series LC pump and a ZORBAX Eclipse XDB C18 column, with distinct mass spectrometric parameters for the parent benzodiazepines and their 7-amino metabolites.

What's inside

  • Abstract
  • Introduction
  • Standards and Reagents
  • Extraction Procedure—Urine
  • Extraction Procedure—Blood
  • Analytical Procedure—Both Urine and Blood
  • Mass Spectrometer Conditions

Specifications

InstrumentAgilent 6410 LC Triple Quadrupole Mass Spectrometer
LC PumpAgilent 1200 Series LC pump
ColumnZORBAX Eclipse XDB C18 4.6 x 50 mm ~ 1.8-um
Operation ModeElectrospray positive mode
Gas Temperature (7-Amino)350°C
Capillary Voltage (7-Amino)4000 V

Frequently asked questions

What drugs are analyzed in this procedure?

The procedure analyzes 14 benzodiazepines and six metabolites simultaneously.

What is the flow rate for the 7-amino metabolites method?

The solvent flow rate for the 7-amino metabolites method is 0.6 mL/min.

What is the column temperature for the 7-amino metabolites method?

The column temperature for the 7-amino metabolites method is 45°C.