Agilent 1100 Series Well-Plate Sampler Optimization Application Note

Optimizing the Agilent 1100 Series well-plate sampler


Agilent 1100 Series Well-Plate Sampler Optimization Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
169 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
16BBBDD815092582F2CE6D32544E3FD3

About this manual

The document details optimization strategies for the Agilent 1100 Series well-plate sampler used in HPLC systems. It focuses on minimizing carry-over through specific needle wash procedures and injector programs that utilize valve bypass positions to clean internal channels.

Additionally, the note covers methods to achieve fast injection cycles by utilizing overlapped injection modes and reducing system delay volume. Performance data regarding linearity, area precision, and cycle times are presented alongside experimental results for various compounds.

What's inside

  • Introduction
  • Instrumentation
  • Optimizing lowest carry-over
  • Fast injection cycles
  • Low delay volume
  • Injector programs
  • Overlapped injection mode
  • Performance data

Frequently asked questions

What is the typical cycle time for drawing and injecting a sample?

The autosampler is able to draw and inject a sample within 36 seconds, including an outside needle wash of 10 seconds.

How does the injector program reduce carry-over?

The injector program switches the injection valve into the bypass position at the end of equilibration to clean the bypass channel and main channel.

How much does switching to overlapped injection mode reduce delay volume?

Switching the injection valve into the bypass position reduces the system delay volume by approximately 300 pl.