Agilent 1100 Series LC/MSD Application Note: Rapid Quantification of an Enkephalin Analgesic in Cerebrospinal Fluid

This document describes the analysis challenges of D-penicillamine enkephalin (DPDPE) and the rapid sensitive analysis of DPDPE in cerebrospinal fluid using an Agilent 1100 Series LC/MSD mass spectrometer. DPDPE is a synthetic enkephalin used for pain control, which has proven difficult to analyze in the past. The document also provides detailed information on the experimental method and sample preparation process.


Agilent 1100 Series LC/MSD Application Note: Rapid Quantification of an Enkephalin Analgesic in Cerebrospinal Fluid - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
90 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
097DDB3BA3ED28B26CBAA6C7646F9481

About this manual

This document presents an application note detailing a method for the rapid quantification of D-penicillamine enkephalin (DPDPE), a synthetic analgesic, in cerebrospinal fluid (CSF). The study, conducted by researchers at the University of California, San Diego in collaboration with Agilent Technologies, addresses the difficulty of analyzing DPDPE due to its disulfide bridge, which makes immunoassays unpredictable.

The note outlines an experimental procedure involving sample extraction using Millipore ZipTip pipette tips followed by analysis via selected ion monitoring (SIM) mode on the Agilent 1100 Series LC/MSD quadrupole mass spectrometer. The document provides specific instrument conditions, including LC and MS parameters, and reports on the method's performance, demonstrating excellent linearity, accuracy within 10.2%, and precision within 9.33% across a concentration range of 1 to 1000 ng/ml.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Instrument Conditions
  • Conclusions
  • References

Specifications

Flow rate0.4 ml/min
Mobile phaseisocratic 75:25 methanol:10 mmol ammonium formate (pH 3)
Ionization modepositive ESI
Nebulizer20 psig
SIM ions646.2 DPDPE, 570.7 IS
Limit of quantitation (LOQ)1.00 ng/ml
Extraction efficienciesbetween 97% and 102%

Frequently asked questions

What is the limit of quantitation for DPDPE analysis?

The limit of quantitation (LOQ) was 1.00 ng/ml.

How accurate and precise is the DPDPE analysis method?

The accuracy was within 10.2% and the precision was within 9.33%.

What is the stability of extracted DPDPE samples?

Extracted samples were stable for 48 hours with a relative standard deviation of +/- 7.4%.