Agilent Multisignal LC/MS Analysis for Compound Screening Application Note

This file describes an LC/MSD autotuning method that can optimize analysis parameters in both positive and negative ionization modes and create a tuning file that contains optimized parameters.


Agilent Multisignal LC/MS Analysis for Compound Screening Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
119 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
E8DF32A1D522E493636CA4EBC5A9DF3F

About this manual

The document describes a technique for acquiring mass spectrometry information using different acquisition modes within a single sample analysis. It highlights how this approach improves productivity in LC/MS compound screening and simplifies method development by allowing the use of a single, generic method.

Specific examples are provided showing the analysis of sulfonamide antibiotics via electrospray ionization (ESI) and polymer additives via atmospheric pressure chemical ionization (APCI). The text details how alternating positive and negative ionization modes maximizes information from each analysis, confirming molecular weights and detecting fragmentation patterns that might be missed in a single mode.

What's inside

  • Introduction
  • Experimental Setup
  • Tuning the LC/MSD
  • ESI-LC/MS Analysis of Sulfonamides
  • APCI-LC/MS Analysis of Polymer Additives
  • Mass Spectra Interpretation

Specifications

System Componentsbinary pump, vacuum degasser, auto-sampler, thermostatted column compartment with column-switching valve, diode-array detector, and an enhanced LC/MSD
Column (Sulfonamides)15 × 3 mm Zorbax SB-C18, 3.5 µm
Flow Rate (Sulfonamides)0.6 ml/min
Column Temperature (Sulfonamides)40°C
Injection Volume (Sulfonamides)5 µl
Drying Gas Flow (ESI)11 l/min
Vcap (Positive Mode, ESI)3000 V
Column Temperature (Polymer Additives)50°C
Flow Rate (Polymer Additives)0.8 ml/min

Frequently asked questions

How does the autotune process work on the LC/MSD?

The autotune process automatically delivers a tune mix, optimizes parameters in both positive and negative ionization modes, and creates a single tune file for polarity switching.

What is the benefit of using positive/negative polarity switching?

It allows confirmation of peak molecular weight without increasing analysis time or sample use by capturing ions in both modes simultaneously.