Agilent Improving the productivity of pharmacokinetic parameter determination Application Note

This Application Note demonstrates how the Agilent LC/MS/MS solution can increase productivity through improved performance of the analytical equipment and superior data analysis software that reports relevant parameters directly from the analytical data. Features of the Agilent solution are: Fast LC methods and short cycle times through alternating column regeneration State-of-the-art MS performance Easy-to-learn and easy-to-use acquisition and data analysis software Calculation of PK parameters directly by the data analysis software Software compliant with the regulations for clinical trials.


Agilent Improving the productivity of pharmacokinetic parameter determination Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
478 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
BDCA41B1604931234876191C1805A46A

About this manual

The document describes a high-throughput analysis method for determining the bio-availability of amoxicillin and clavulanic acid. It utilizes an Agilent 1200 Series Rapid Resolution LC system coupled with an Agilent 6410 triple quadrupole mass spectrometer to achieve cycle times of less than 1.5 minutes.

What's inside

  • Introduction
  • Equipment
  • LC method
  • MS method
  • Plasma work-up
  • MRM settings
  • Results and discussion
  • Method development
  • Data Analysis

Specifications

Flow rate0.8 mL/min
Column temperature40°C
Injection volume4 µL
Total run time1.3 min
Amoxicillin calibration range10 - 10000 ng/mL
Clavulanic acid calibration range50 - 5000 ng/mL

Frequently asked questions

What is the cycle time of the described LC method?

The cycle time was below 1.5 minutes.

Which software is used for data analysis and PK parameter calculation?

Agilent MassHunter workstation software is used to calculate pharmacokinetic values directly.

What are the MRM settings for Amoxicillin?

Amoxicillin is analyzed in positive mode with a precursor ion of 366.2, quantifier ion 114, and qualifier ion 208.