Agilent 6510 Q-TOF LC/MS Application Note

This study examines the effects of precursor and fragment ion mass accuracy on protein identification using the Accurate Mass Q-TOF. Results demonstrate that fragment ion mass accuracy is as important (if not more important) than precursor ion mass accuracy in reducing false positive identification rates in MS/MS database searches, and plays a critical role in improving the throughput of proteomics research projects.


Agilent 6510 Q-TOF LC/MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
469 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
0CEA1C2734ACB0C57D149E7AD28A9E9E

About this manual

This document is an application note from Agilent Technologies that investigates the effect of fragment ion mass accuracy on protein identification rates in shotgun proteomics. It compares the performance of the Agilent 6510 Q-TOF LC/MS system against an Agilent 6330 Ion Trap mass spectrometer using HeLa cell samples.

The study demonstrates that high fragment ion mass accuracy is as important as precursor ion mass accuracy in reducing false positive identification rates. The authors utilized the 6510 Q-TOF with specific mass tolerances (10 ppm for precursors and 40 ppm for fragments) to achieve higher identification yields and confidence compared to the ion trap system.

What's inside

  • Abstract
  • Introduction
  • Materials and Methods
  • Hela Cell Culture
  • OFFGEL Fractionation
  • LC/MS Analysis
  • Database Searches
  • Results and Discussion
  • Selection of MMD values
  • Comparison of ion trap and Q-TOF results

Specifications

Acquisition rate20 spectra/sec
Resolving power15,000
Dynamic range3.5 orders of magnitude
Precursor ion tolerance (Q-TOF)10 ppm
Fragment ion tolerance (Q-TOF)40 ppm
Precursor ion tolerance (Ion Trap)2.5 Da
Fragment ion tolerance (Ion Trap)0.7 Da
MS Scan range (Q-TOF)300-2000 m/z
MS/MS Scan range (Q-TOF)50-3000 m/z
MS Scan range (Ion Trap)300-1800 m/z
MS/MS Scan range (Ion Trap)100-2000 m/z

Frequently asked questions

What is the optimal precursor ion tolerance for the 6510 Q-TOF?

The optimal precursor ion tolerance for the 6510 Q-TOF is 10 ppm.

What is the optimal fragment ion tolerance for the 6510 Q-TOF?

The optimal fragment ion tolerance for the 6510 Q-TOF is 40 ppm.

Does fragment ion mass accuracy improve peptide identification?

Yes, high fragment ion mass accuracy is as important as precursor ion mass accuracy in reducing false positive identification rates.