Agilent 6890/5973N GC/MSD Pesticide Screening Application Note

Agilent Technologies' new fast GC/MSD method can significantly speed up pesticide screening


Agilent 6890/5973N GC/MSD Pesticide Screening Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
81 KB
Published
12 June, 2012
Updated
01 October, 2023
Model
5973N
MD5 checksum
B175DF84EB66911509F8A4172916C348

About this manual

Agilent Technologies developed a new, fast GC/MSD method to significantly speed up the screening of pesticides and endocrine disrupters. The method utilizes Agilent's Method Translation (MTL) and Retention Time Locking (RTL) software to scale a standard 42-minute analysis down to as little as 10.5 minutes without losing resolution. A specific combination of column stationary phase, carrier gas flow rate, and oven temperature programming is required to lock all compounds to an expected retention timetable.

The document details the experimental setup, including the use of a 10 m x 0.1 mm x 0.1 µm HP-5 column and a 16:1 split ratio. It presents chromatographic conditions for onefold, twofold, threefold, and fourfold speed gains in a table. The results show that the scaled screening libraries can accurately correspond to the faster analyses, making the screening process more powerful and adaptable.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • General chromatographic conditions
  • Table 1. Chromatographic Conditions
  • Figure 2. Three TICs
  • Conclusion
  • References

Specifications

Analysis time (standard)42 minutes
Minimum analysis time10.5 minutes
Column length10 m
Column inner diameter0.1 mm
Column film thickness0.1 µm
Split ratio16:1
Carrier gasHelium
Injector temperature250°C
Source temperature230°C
Transfer line temperature280°C
High mass500 amu

Frequently asked questions

What software is used to develop the new fast GC/MSD method?

Agilent's GC method translation software (MTL) and retention time locking (RTL) software were used.

How many common pesticides and endocrine disrupters are screened?

The method screens for 567 of the most common pesticides and endocrine disrupters.

What column was used to increase analysis speed?

A 10 m x 0.1 mm x 0.1 µm HP-5 column was used.