Detection of Low Levels of Carbaryl in Food Using Agilent LC/MSD Trap System Application Note

This application note describes the application of ion trap LC/MS/MS to determine pesticide and herbicide contamination in food samples. This method uses ion trap LC/MS/MS to confirm, identify and quantify low levels of carbaryl in foods. This is mainly due to the high sensitivity and specificity of MS/MS with the ion trap.


Detection of Low Levels of Carbaryl in Food Using Agilent LC/MSD Trap System Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
370 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
FC77BC5ED7C12484DB31E32E6EFEE295

About this manual

The document details an analytical method developed to determine the presence of the carbamate pesticide carbaryl in composite food extracts. The study addresses limitations of previous fluorescence techniques, such as false positives and insufficient sensitivity, by utilizing LC/MS/MS with ion trap technology for absolute confirmation and quantification.

Experimental procedures involved a nine-step extraction scheme using acid precipitation, Soxhlet extraction, and cleanup via GPC. The method achieved a limit of detection (LOD) of 1 ng/g and a limit of quantitation (LOQ) of 10 ng/g in food matrices, demonstrating significantly higher sensitivity compared to HPLC fluorescence methods.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Conclusions

Frequently asked questions

What is the limit of detection for carbaryl in food using this method?

The limit of detection in the food matrix was found to be 1 ng/g.

What is the limit of quantitation used for determining carbaryl in food?

The limit of quantitation used for the determination in food was 10 ng/g.

What is the concentration range over which the calibration curve is linear?

The calibration curve based on the m/z 145 product ion is linear over the concentration range of 1-1000 ng/g.