Agilent 2100 Bioanalyzer Application Note: Meat Speciation Assays

This document describes the application of the Agilent 2100 bioanalyzer with the DNA 500 LabChip® kit for the development of meat speciation assays. By using the bioanalyzer to confirm the suitability of primer sets before probe production, the cost of probe production can be reduced.


Agilent 2100 Bioanalyzer Application Note: Meat Speciation Assays - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
119 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
6EDD401114B52878A9263A7E16006430

About this manual

This document details a method for developing real-time PCR assays for food authenticity, specifically for detecting and quantifying meat species. It highlights the use of the Agilent 2100 Bioanalyzer to verify that newly designed PCR primers function correctly and are specific before the costly production of fluorescent probes is undertaken. The authors demonstrate how this approach saves expenses by identifying non-specific primers early in the development process.

What's inside

  • Abstract
  • Introduction
  • Materials and Methods
  • Design of PCR assay
  • Performance of PCR reaction
  • DNA 500 LabChip preparation
  • Results and Discussion
  • Species-specific assay development
  • Total meat assay development
  • Conclusion

Frequently asked questions

What is the purpose of using the Agilent 2100 bioanalyzer with the DNA 500 LabChip kit in this application?

The kit is used to confirm that newly designed PCR primers are functioning correctly before the more expensive fluorescent probes are produced.

What PCR reaction volume and primer concentration were used for the performance tests?

The PCR reactions were performed in 25-11 volumes using 300 nM of each primer.

What are the specific temperature and time settings for the TaqMan-based amplification protocol?

The protocol consists of 30 cycles of a two-step reaction at 95 °C for 15 seconds and 60 °C for one minute, followed by a final 10-minute step at 72 °C.