Agilent Pesticide Identification Quantitation Using Retention Time Locking GC/MS Application Note

This document describes a method for quantifying pesticides in the parts-per-trillion range using retention time locking and GC/MS application. By using a selected ion monitoring method and large volume injection capability, the target compound quantitation limit can be lowered to the low parts-per-billion (pg/µL) range. Additionally, an automated 567-compound retention time locking pesticide mass spectral library can screen an acquired sample's data file for all 567 compounds in seconds. The library can also be used for rapid screening of samples acquired in selected ion monitoring method. Using the compound library information, a selected ion monitoring method for 80 target compounds was created in less than 2 hours without running any analyses.


Agilent Pesticide Identification Quantitation Using Retention Time Locking GC/MS Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
471 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
3D7D69312EC1D947148365BADF628202

About this manual

The document details an analytical approach to lower mass spectrometer detection limits from sub-ppm to low parts-per-billion or parts-per-trillion levels. This is achieved by combining selected ion monitoring (SIM) mode, large volume injections (LVI), and higher electron multiplier voltage with a specialized retention time locking (RTL) pesticide library.

The text provides specific system configurations for the Agilent 6890 GC and 5973 MSD, including column specifications and software versions. It outlines experimental procedures using an 80-pesticide mixture and explains how to construct SIM methods from the RTL database to improve signal-to-noise ratios and simplify data analysis.

What's inside

  • Application
  • Abstract
  • Introduction
  • System Configuration
  • GC Method Parameters
  • MS Method Parameters
  • Injection Parameters
  • Results and Discussion
  • RTL Database Usage

Frequently asked questions

What is the typical quantitation limit for a mass spectrometer in Scan mode?

The typical pesticide quantitation limit for a mass spectrometer in the Scan mode is in the sub-ppm range.

How can target compounds at parts-per-trillion be quantified?

Target compounds at parts-per-trillion can be quantified by adding large volume injection capability to a retention time locked gas chromatography/mass spectrometry method.

What software version is required for the MSD ChemStation?

G1701BA version B.00.00 MSD ChemStation software or higher is required.