Agilent 7500c ICP-MS Measurement of Macro Trace Elements in Plant Digests Application Note

This application note describes the use of the Agilent 7500c ICP-MS system to measure major, minor and trace elements in two standard reference plant materials.


Agilent 7500c ICP-MS Measurement of Macro Trace Elements in Plant Digests Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
385 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
7500c
MD5 checksum
B9298FF998254370A68BBEDE9CAE23C8

About this manual

The document describes the use of the Agilent 7500c ICP-MS system equipped with an Octopole Reaction System (ORS) for analyzing major, minor, and trace elements in standard reference plant materials. The study compares results obtained with the 7500c against certified reference values and data from inductively coupled plasma optical emission spectroscopy (ICP-OES). The analysis confirms that the 7500c provides accurate results for all elements, effectively handling complex sample matrices.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • Instrument Tuning
  • Results and Discussion
  • Operating Conditions

Specifications

Plasma RF power1500 W
Sample flow rate240 pL/min
Spray chamber temperature2°C
Carrier gas flow1.1 L/min
Sample depth9.5 mm from load coil

Frequently asked questions

What elements were analyzed in the plant digests?

The analysis covered major, minor, and trace elements including Potassium, Calcium, Chromium, Iron, Copper, Zinc, Arsenic, Selenium, and Cadmium.

How does the 7500c handle interferences?

The system uses an Octopole Reaction System (ORS) to remove spectral interferences by reacting a gas with the interference or using energy discrimination.

Download PDF (385 KB)