Agilent 6890/5973N GC/MSD Fast Screening Application Note Part II

Agilent Technologies' new, fast GC/MSD method can significantly speed up the screening of pesticides. Agilent's GC Method Translation software was used in developing the new method based on the standard 42-min method. A 15 m × 0.25 mm × 0.25 mmmmm Agilent HP-5MS column was used to increase analysis speed up to fourfold. The time savings were implemented in increments (down to 10.5 minutes) to verify the predictability of scaling and the effect of scaling on the signal-to-noise ratio.


Agilent 6890/5973N GC/MSD Fast Screening Application Note Part II - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
82 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
5973N
MD5 checksum
80D93479B4C78BA555F202DDCFC1AB3B

About this manual

Agilent Technologies developed new techniques to speed up the screening of pesticides and endocrine disrupters. This document details a method that reduces analysis time from 42 minutes down to as little as 10.5 minutes using specific column dimensions and optimized temperature programming.

The approach utilizes Agilent's GC Method Translation (MTL) software to adjust operating conditions for faster runs, followed by Retention Time Locking (RTL) to ensure accurate compound identification despite the reduced run times. The document provides experimental data comparing standard, 2.5x speedup, and 4x speedup methods.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Chromatographic Conditions
  • MS Conditions
  • Conclusion

Frequently asked questions

What is the minimum run time for this fast screening method?

The run time can be shortened to 10.5 minutes using a fast oven ramp rate and a specific column.

Which Agilent GC option is required for the faster oven ramp used in this study?

The Agilent 6890 GC fast oven option (220/240V in the U.S.) is required.

What carrier gas flow rate can a turbo pump handle for this method?

A turbo pump used in this study could handle a 3.8 mL/min carrier flow.