Agilent System suitability testing for Tramadol quality control with the Agilent 1120 Compact LC ZORBAX C-18 columns Application Note

The document describes the system suitability testing for Tramadol quality control using the Agilent 1120 Compact LC and ZORBAX C-18 columns. The system demonstrates excellent retention time precision and peak area precision, making it suitable for drug analysis.


Agilent System suitability testing for Tramadol quality control with the Agilent 1120 Compact LC ZORBAX C-18 columns Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
305 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
760DD70A6EA4ECBF4881B00873F70981

About this manual

The document describes an analytical method to evaluate the performance of the Agilent 1120 Compact LC system. It focuses on system suitability testing (SST) for the quantification of the analgesic drug Tramadol and potential synthesis byproducts, including starting materials like 3-bromoanisole.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Sample preparation
  • Chromatographic parameters
  • Sequence table
  • Results and discussion
  • Acceptance criteria

Specifications

Column dimensions250 x 4.6 mm
Particle size5 µm
Flow rate1.2 mL/min
Injection volume10 µL
Column oven temperature30 °C
Detection wavelength270 nm

Frequently asked questions

What are the acceptance criteria for retention time precision?

The relative standard deviation (RSD) of retention time must be less than 0.07 %.

What is the required resolution limit for system suitability testing?

The resolution between peaks must be greater than 2.00.

How many theoretical plates are required for the column efficiency test?

The system must demonstrate more than 2000 theoretical plates.

Download PDF (305 KB)