Agilent 5990-3528EN- Running fast LC within USP limits Case study with USP pravastatin sodium chromatographic purity method using the Agilent 1200 Series Rapid Resolution LC system Application Note

Using the Agilent 1200 Series Rapid Resolution LC system and Agilent Method Translator software, USP pravastatin sodium chromatographic purity method is used to conduct a fast LC analysis. This application note describes how to start from a USP method, incorporate the allowed adjustments, and produce a fast method that meets system suitability requirements.


Agilent 5990-3528EN- Running fast LC within USP limits Case study with USP pravastatin sodium chromatographic purity method using the Agilent 1200 Series Rapid Resolution LC system Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
292 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
F1532CBF2EBF0C33A883F7A7D8BE7A7A

About this manual

Recent revisions in United States Pharmacopeia (USP) general chapter <621> allow for adjustments to chromatographic parameters to enhance chromatogram quality and reduce run times. This document details a case study where a standard 30-minute USP method was optimized into a faster method using the Agilent 1200 Series Rapid Resolution LC (RRLC) system.

The study utilizes Agilent Method Translator software to adjust parameters such as particle size, flow rate, column dimensions, and temperature. The optimized method uses a 1.8 µm particle column and a 1.5 mL/min flow rate, significantly reducing analysis time while maintaining system suitability requirements for pravastatin sodium and its impurities.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Equipment
  • Parameters
  • Results and discussion
  • System suitability
  • Linearity test

Specifications

Column dimensions (Fast method)4.6 mm x 30 mm
Column particle size (Fast method)1.8 µm
Flow rate (Fast method)1.5 mL/min
Column temperature (Fast method)35 °C
Wavelength for DAD238 nm
SystemAgilent 1200 Series Rapid Resolution system

Frequently asked questions

How is the fast method developed from the USP method?

The fast method is developed by using Agilent Method Translator software to incorporate allowed USP adjustments such as smaller particle size and higher flow rates.

What are the permitted adjustments for the USP pravastatin method?

Permitted adjustments include column length, column inner diameter, particle size, flow rate, injection volume, and column temperature.

Does the fast method meet system suitability requirements?

Yes, the fast method meets requirements with an RCA RRT of about 1.1 and a resolution of 3.8.