Generating nonlinear calibration curve using ChemStation software for analysis of topiramate with Agilent 1200 Series Ev

This application note describes the method for generating non-linear ELSD calibration curves using ChemStation software and the quantification of non-UV active compound topiramate using this curve.


Generating nonlinear calibration curve using ChemStation software for analysis of topiramate with Agilent 1200 Series Ev - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
154 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
F5E068A9F73E2A5E987376F445A83B24

About this manual

Evaporative light scattering detection (ELSD) is useful for detecting and quantifying non-UV active compounds that are semivolatile or nonvolatile. The response of an ELSD to concentration is typically nonlinear, often following an exponential relationship described by the equation y = ax^b.

This study utilized Agilent ChemStation software to generate a power curve calibration, which matches the typical ELSD response equation. Topiramate was analyzed using an Agilent 1200 Series Evaporative Light Scattering Detector with an Agilent ZORBAX Eclipse XDB phenyl column.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • Results and discussion
  • Conclusion

Frequently asked questions

What concentration range was used for the nonlinear calibration curve?

The curve was plotted using external standards in the range of 10.3 pg/mL to 137.7 pg/mL.

What was the average value and standard deviation for repeated injections of the same solution?

Repeated injections gave an average value of 26.2 mg with a standard deviation of 1.2.

What are the dimensions of the Agilent ZORBAX Eclipse XDB phenyl column used?

The column used was 50 mm x 3.0 mm with a particle size of 1.8 pm.