Agilent System suitability testing for Aripiprazole quality control with the Agilent 1120 Compact LC ZORBAX C-18 columns Application Note

This Application Note describes the System Suitability Testing (SST) results of Aripiprazole quality control with the Agilent 1120 Compact LC and ZORBAX C-18 columns. The system is shown to have excellent precision in retention time, area and height measurements.


Agilent System suitability testing for Aripiprazole quality control with the Agilent 1120 Compact LC ZORBAX C-18 columns Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
296 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
AC31A51E9473036BF844917755E28658

About this manual

This Application Note evaluates the performance of the Agilent 1120 Compact LC system for the analysis of Aripiprazole and its precursor impurities. The study focuses on System Suitability Testing (SST) to verify instrument precision and accuracy for pharmaceutical quality control applications.

The document details the experimental setup, including the use of an Agilent ZORBAX Eclipse XDB C18 column and Agilent EZChrome Elite Compact Compliance software. It presents results from 10 replicate injections, confirming that the system meets stringent acceptance criteria for retention time, peak area, resolution, and theoretical plates.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Compounds
  • Preparation of samples
  • Chromatographic parameters
  • Sequence table
  • Results and discussion
  • Acceptance criteria
  • System suitability test results

Specifications

ColumnAgilent ZORBAX Eclipse XDB C18, 5µm, 150 x 4.6 mm
Flow rate1.0 mL/min
Column oven temperature40 °C
Detection wavelength254 nm
Injection volume10 µL
Total run time9 min

Frequently asked questions

What are the acceptance criteria for retention time precision?

The relative standard deviation (RSD) of retention time must be less than 0.07 %.

What is the limit for peak area precision?

The RSD of area must be less than 1.00 %.

What is the minimum required resolution?

The resolution must be greater than 2.00.