Agilent 325-MS Triple Quadrupole Mass Spectrometer Application Note 02219

This application note describes a new EPA drinking water LC/MS/MS method for the determination of non-volatile compounds on the Varian 325-MS triple quadrupole mass spectrometer.


Agilent 325-MS Triple Quadrupole Mass Spectrometer Application Note 02219 - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
7
File format
PDF
File size
259 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
325-MS
MD5 checksum
1D5878BA71CA817F73367CB0E1E3336C

About this manual

This application note evaluates a new U.S. EPA method for determining semi-volatile and non-volatile contaminants in drinking water using the Varian (now Agilent) 325-MS triple quadrupole mass spectrometer equipped with a vortex ESI (vESI™) source.

The study details the optimization of LC/MS/MS conditions, including HPLC parameters and MS settings such as gas temperatures and pressures. It demonstrates how programming these parameters using SelecFlow™ and SelecTemp™ features improves sensitivity for various analytes compared to fixed temperature settings.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions
  • MS Parameters
  • Method Optimization
  • Results and Discussion

Frequently asked questions

What source is used with the 325-MS in this method?

The vortex ESI (vESI™) source is used together with the ability to program temperature and flow rate of atmospheric pressure interface gases.

What are the acceptance criteria for calibration recovery?

The lowest calibration point must have +50% recovery, and all other levels must have +30% recovery.