Agilent Analysis of (±)-11-Nor-9-Carboxy-Delta-9-THC in Urine by Negative Ion Electrospray LC/MS/MS Application Note

This document describes a simple, high-throughput LC/MS/MS method for the detection and quantitation of (±)-11-nor-9-carboxy-delta-9-THC in urine using negative ion electrospray. The method does not require derivatization and provides the same specificity and sensitivity as GC/MS.


Agilent Analysis of (±)-11-Nor-9-Carboxy-Delta-9-THC in Urine by Negative Ion Electrospray LC/MS/MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
293 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
1200L
MD5 checksum
7E78F556563DDA6F522A572B15976B88

About this manual

The document details an analytical method using the Varian (now Agilent) 1200L LC/MS/MS system equipped with an ESI source to analyze urine samples for the THC metabolite (+)-11-nor-9-carboxy-delta-9-THC. The method utilizes negative ion electrospray ionization and eliminates the need for time-consuming sample derivatization required by traditional GC/MS techniques.

Specific instrumentation and conditions are provided, including the Varian ProStar 410 AutoSampler, ProStar 210 Solvent Delivery Modules, and a Varian Pursuit Diphenyl column. The method achieves a linear range from 1 ppb to 1000 ppb with a retention time of 4.6 minutes for the analyte.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions
  • MS Parameters
  • Scan Parameters
  • Results and Discussion
  • Conclusion

Specifications

ColumnVarian Pursuit Diphenyl 3 um, 50 x 2 mm
Linear Range1 ppb to 1000 ppb
Retention Time4.6 minutes

Frequently asked questions

What is the linear range of the LC/MS/MS method for carboxy-THC?

The method is linear from 1 ppb to 1000 ppb.

Does this method require sample derivatization?

No, the LC/MS/MS method provides specificity and sensitivity without the need for a derivatization process.

What is the retention time for the THCC analyte?

The retention time for the THCC analyte is 4.6 minutes.

Download PDF (293 KB)

More Agilent Manuals