Agilent 1200L LC/MS/MS Application Note for Dinitro Flavonoid Quantitation

This document describes the quantitation of a dinitro flavonoid (DNF) in rat plasma using the Varian 1200L LC/MS/MS system. Flavonoids are compounds that make up the pigments in fruits, vegetables, and herbs. There have been reports of flavonoids having multiple beneficial therapeutic effects for a large number of serious illnesses, such as rheumatoid arthritis, cancer, cataracts, anxiety, and depression. The objective of this work was to carry out a pharmacokinetic evaluation to determine a profile with the formulation being tested at increasing doses where there is pharmacological effect. A Varian 1200L LC/MS/MS system was used for the quantitation of plasma levels of a novel dinitro-substituted flavone (DNF) in samples from a pharmacokinetic study following oral administration of the drug. Flavone was used as the internal standard. Structures are shown in Figure 1.


Agilent 1200L LC/MS/MS Application Note for Dinitro Flavonoid Quantitation - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
307 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
3E9DFACF2076CF311DD64EF16B71BC15

About this manual

This document details a pharmacokinetic evaluation method for determining the profile of a novel dinitro-substituted flavone (DNF) in rat plasma. The study utilized oral administration and analyzed samples to observe concentration changes over time, noting phenomena such as entero-hepatic recirculation.

The analysis was performed using an Agilent 1200L LC/MS/MS system equipped with an ESI source. The method involves extracting DNF from plasma using hexane:chloroform and analyzing it via Selected Reaction Monitoring (SRM). Specific HPLC conditions, MS parameters, and scan settings for the analyte and internal standard are provided.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions
  • MS Parameters
  • Scan Parameters
  • Results and Discussion

Frequently asked questions

What system was used for the quantitation of DNF in this study?

A Varian (now Agilent) 1200L LC/MS/MS system equipped with an ESI source.

What internal standard was used in the analysis?

Flavone was used as the internal standard.

What is the calibration curve range for DNF in rat plasma?

The calibration curve range is from 0.10 ng/mL to 5.00 ng/mL of rat plasma.