Agilent SI-01375 High Productivity Analysis for Pesticide Residues in Vegetables Application Note

This document describes a method for pesticide residue analysis in agricultural products. By using a 0.15 mm small diameter capillary GC column, 200 pesticide residues can be separated and quantified quickly and successfully. This method improves instrument productivity and reduces detection limits while improving signal to noise ratios.


Agilent SI-01375 High Productivity Analysis for Pesticide Residues in Vegetables Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
117 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
69CE010E7342E4087EDFD23CF1B8C940

About this manual

This document presents a method for the high productivity analysis of pesticide residues in vegetable matrices using Gas Chromatography/Mass Spectrometry (GC/MS). The primary focus is on utilizing 0.15 mm small diameter capillary GC columns to significantly reduce analysis time while maintaining separation quality.

The note details specific instrument conditions, including the use of a Varian CP-3800/450-GC and a Varian 4000/240-MS Ion Trap detector. It highlights that the method allows for the separation and quantification of 200 pesticides in less than 20 minutes, compared to 35 minutes with standard columns, while increasing sensitivity four-fold.

What's inside

  • Introduction
  • Chromatogram
  • GC Conditions
  • Analysis Time
  • Sensitivity
  • PTV Injection

Specifications

Column ID0.15 mm
Column Film Thickness0.15 um
Carrier GasHelium
Injector TypePTV
DetectorIon Trap
Analysis Time (New)less than 20 min
Analysis Time (Original)35 min

Frequently asked questions

What is the main benefit of using 0.15 mm small diameter capillary GC columns?

The main benefit is a significant reduction in analysis time, increasing instrument productivity by as much as 50% in some cases.

How does the sensitivity change when using the 0.15 mm ID column?

Overall sensitivity has increased four fold due to fast elution on the 0.15 mm ID column.

Which injector is recommended to reduce thermal degradation of thermally labile pesticides?

A temperature programmable splitless injection using the Varian 1079 Programmable Temperature Vaporizing Injector (PTV) is recommended.