Varian PIONA+ Analyzer Fast Analysis Application Note

This application note describes work to determine if concurrent heating could be used to significantly reduce the analysis time and improve sample throughput in the PIONA+ mode.


Varian PIONA+ Analyzer Fast Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
446 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
4FFDEC06B0DB39D5860FD35A12C9C907

About this manual

This document details a method to significantly reduce the analysis time for characterizing hydrocarbon samples using the Varian PIONA+ Analyzer. The standard analysis time is approximately 180 minutes, which limits daily sample throughput. The note introduces a technique called "FastPIONA" mode.

What's inside

  • Introduction
  • Materials and Reagents
  • Instrumentation
  • Results and Discussion
  • FastPIONA Mode
  • Conclusion

Specifications

Standard Analysis Time180 minutes
Fast Analysis Time95 minutes
Analysis Reduction~50 % less
Compliance StandardBS.EN 14517

Frequently asked questions

How much does concurrent heating reduce analysis time?

It reduces total analysis time from 180 minutes to 95 minutes, which is approximately 50% less.

Does the FastPIONA mode require hardware changes?

No, the reduction in analysis time is possible without any hardware changes.

Does the fast mode affect separation quality?

No, the elution integrity of the component groups remains intact with no negative influence on separation quality.