Agilent Low Level Measurement of Cadmium in Foods Application Note

This method is applicable to a wide range of animal and plant tissues including dairy products, grains, fruit and vegetables and oil seeds such as poppy seeds. It is suitable for cadmium concentration in the range 0.01-0.3 µg/g. Sample is digested by wet ashing and cadmium is then complexed and extracted into an organic solvent. After removal of the spent aqueous phase the organo metallic complex is destroyed by the addition of dilute HNO3. This aqueous phase containing the stripped cadmium is then analyzed by Graphite Furnace Atomic Absorption Spectroscopy (GFAAS). The additional stripping step is employed to improve precision and facilitate the use of aqueous standards.


Agilent Low Level Measurement of Cadmium in Foods Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
86 KB
Published
12 June, 2012
Updated
30 September, 2023
Models
GTA-95, AA-875
MD5 checksum
9B7DA5FA03814C2EA09C6B1BE9CA3766

About this manual

This document outlines a procedure for determining cadmium concentrations in animal and plant tissues, including dairy products, grains, fruits, vegetables, and oil seeds. The method is suitable for cadmium concentrations in the range of 0.01-0.3 µg/g. The process involves digesting samples via wet ashing, complexing cadmium with ammonium pyrrolidine dithiocarbamate (APDC), and extracting it into di-isobutyl ketone (DIBK). The organic phase is then stripped with dilute nitric acid, and the resulting aqueous phase is analyzed using Graphite Furnace Atomic Absorption Spectroscopy (GFAAS).

What's inside

  • Reagents and Chemicals
  • Instrumentation
  • Sample Preparation
  • Procedure
  • Results
  • Accuracy
  • Detection Limit
  • Precision
  • Quality Assurance

Specifications

Applicable Cadmium Concentration Range0.01-0.3 µg/g
Instrumentation UsedAgilent GTA 95 with Agilent AA-875
Detection Limit (Solution)Better than 0.0005 mg/mL Cd
Reporting Limit (Grains)0.02 mg Cd/g

Frequently asked questions

What is the suitable range for cadmium concentration in this method?

The method is suitable for cadmium concentration in the range 0.01-0.3 µg/g.

Which instruments are used for the analysis?

An Agilent GTA 95 graphite furnace atomizer with autosampler is used with an Agilent AA-875.

What is the detection limit for cadmium in solution?

The detection limit is better than 0.0005 mg/mL Cd in solution.

What is the recommended recovery range for quality assurance?

The recovery should be between 80% and 120%.