Agilent 320-MS Triple Quadrupole LC/MS/MS Application Note for Imazaquin Analysis

This document describes a method for quantitative analysis of the herbicide imazaquin on the Varian 320-MS Triple Quadrupole LC/MS/MS. The method is highly sensitive, selective and accurate, and can detect imazaquin at trace levels.


Agilent 320-MS Triple Quadrupole LC/MS/MS Application Note for Imazaquin Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
72 KB
Published
12 June, 2012
Updated
01 October, 2023
Model
320-MS
MD5 checksum
AE8C1862515306EAADBFB8A334785D5C

About this manual

The document details a method developed to quantify the concentration of the imidazolinone herbicide imazaquin at trace levels. The analysis utilizes an ESI positive ionization mode and specific HPLC conditions, including a Polaris C18-A column and a mobile phase of 60:40 water:methanol with 0.1% formic acid.

What's inside

  • Introduction
  • Sample Preparation
  • HPLC Conditions
  • MS Parameters
  • Results & Discussion

Frequently asked questions

What is the detection range for imazaquin using this method?

The method can detect and quantify imazaquin from 0.05 ppb to 200 ppb.

What is the correlation coefficient for the calibration curve?

The calibration curve for imazaquin has an r value of 0.999.

What is the signal-to-noise ratio for a 20-pL injection of a 0.05 ppb standard?

An RMS signal-to-noise ratio of 61 was obtained for a 20-pL injection of a 0.05 ppb standard.