Agilent Saturn 2000 Identification of Essential Oil Components by EI/MS Application Note

This application note describes a method for identifying essential oil components using EI/MS and mixed reagent CI/MS. This method can improve the qualitative identification of essential oil components.


Agilent Saturn 2000 Identification of Essential Oil Components by EI/MS Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
122 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
1360892ECC64826CDCF094055BFB08C5

About this manual

This application note presents a strategy to improve the qualitative identification of essential oil components using Gas Chromatography/Mass Spectrometry (GC/MS). The method combines Electron Impact (EI) mass spectrometry with a commercially available flavor and fragrance library and a secondary confirmation technique using Chemical Ionization (CI) with a 50:50 mix of acetonitrile and d3-acetonitrile.

What's inside

  • Introduction
  • Experimental
  • GC and MS Conditions
  • Results and Discussion
  • Conclusions
  • Reference

Specifications

ColumnDB5 30m x 0.25 mm ID x 0.25um film
Injection Volume1 wL
Split Ratio20:1
Carrier Gas Flow1 mL/min
Initial Temperature60°C
Final Temperature240°C
Ramp Rate3°C/min
MS AGC Target10,000
Scan Range EI40-500u
Scan Range CI90-500u
Scan Rate1.0 sec/scan

Frequently asked questions

What reagents are used for the dual CI confirmation?

A 50:50 mix of acetonitrile and d3-acetonitrile is used as the chemical ionization reagent.

How is the essential oil sample prepared to avoid overloading?

The essential oils are diluted 50:1 in hexane and then split 20:1 in the injector.

What library is recommended for the EI/MS search?

The Adams Flavor and Fragrance Library is recommended, as it contains retention time data to improve the search process.