Agilent GCMS49 Analysis of Hydrolysed Vegetable Protein for Chloropropandiols Using Selected Ion Storage Application Not

This document describes a method for the analysis of chloropropanols in hydrolyzed vegetable protein using selected ion storage. The method is based on the capillary GC of a heptafluorobutyrate derivative with electron capture detection (ECD) at an overall limit of detection of about 1.5 mg/kg, depending on the sample type (dry weight basis).


Agilent GCMS49 Analysis of Hydrolysed Vegetable Protein for Chloropropandiols Using Selected Ion Storage Application Not - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
110 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
8A5854CE25365CCE7826DC81B83CACC0

About this manual

This document details an analytical method for detecting chloropropanols, specifically 3-monochloropropandiol (3-MCPD), in Hydrolysed Vegetable Protein (HVP). The method utilizes Selected Ion Storage (SIS) mode on a GC/MS system to overcome matrix interference and improve detection limits compared to traditional Electron Capture Detection (ECD).

The note provides the full experimental procedure, including sample preparation involving extraction with hexane/ether and derivatization using heptafluorobutyrylimidazole. It also lists specific instrument conditions for the gas chromatograph and mass spectrometer required to achieve a detection limit of approximately 0.01 mg/kg.

What's inside

  • Discussion
  • Figure 1A: Full scan spectrum
  • Figure 1B: Spectrum using Selected Ion Storage
  • Experimental Procedure
  • Instrument Conditions
  • Table 1: Recovery data
  • Conclusion

Specifications

Detection Limit0.01 mg/kg
Column TypeBPX5 25M X 0.2mm X 0.25um
Mass Range70-300u
Internal Standardp-dichlorobenzene

Frequently asked questions

What is the detection limit for 3-MCPD in this method?

The limit of detection is about 0.01 mg/kg based on a spike of 0.05 mg/kg.

What internal standard is used for quantitation?

Quantitation is performed using p-dichlorobenzene as the internal standard.

How much does Selected Ion Storage improve detection compared to ECD?

Selected Ion Storage provides a 150-fold increase in detection sensitivity over existing procedures.