Agilent 490 Micro GC Application Note: Permanent Gases on a COX Module

This application note demonstrates the capabilities of the Agilent J&W CP-COX column with the Agilent 490 Micro GC, including separation of permanent gases and backflush possibilities to ensure extended column lifetimes.


Agilent 490 Micro GC Application Note: Permanent Gases on a COX Module - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
610 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
0DD077D7518AF0D74A43CEDC741DC198

About this manual

This document describes an application using the Agilent J&W CP-COX column with an Agilent 490 Micro GC system to analyze permanent gases. It highlights that while traditional Molsieve columns require high-temperature bakeouts to regenerate, the CP-COX column allows carbon dioxide to elute and offers a prolonged lifetime. The note details how a precolumn with backflush can remove water and higher hydrocarbons to prevent column pollution.

The document presents experimental results showing the repeatability of the analysis for components including helium, hydrogen, oxygen, nitrogen, carbon monoxide, methane, and carbon dioxide. It concludes that the CP-COX column is a good alternative to Molsieve for permanent gas analysis, ensuring reliable results with good repeatability figures.

What's inside

  • Abstract
  • Introduction
  • Instrumentation
  • Results and Discussion
  • Conclusion
  • For More Information

Specifications

Column Temperature100 °C
Backflush to Vent Time13s
Injection Time80 ms
Injection Temperature110 °C
Sample Line Temperature100 °C
Sampling Time30s
Stabilization Time5s
Run Time200s

Frequently asked questions

What are the repeatability figures for the analyzed components?

The repeatability figures per component on peak area are: He (0.43%), H2 (0.47%), N2+O2 (0.47%), CO (0.64%), CH4 (0.33%), and CO2 (0.51%).

Does the CP-COX column separate oxygen and nitrogen?

No, the CP-COX column does not separate oxygen and nitrogen; they elute as a single peak.

What happens to water and higher hydrocarbons during analysis?

Water and higher hydrocarbons are backflushed to the vent to prevent them from entering the analytical column and polluting it.