Agilent Bond Elut Plexa Application Note Reduced Ion Suppression Improved LC/MS Sensitivity

Agilent Bond Elut Plexa polymeric SPE reduces ion suppression for better sensitivity and low LOD and LOQ. Beta blockers are extracted with excellent correlation coefficients, recoveries, and % RSD.


Agilent Bond Elut Plexa Application Note Reduced Ion Suppression Improved LC/MS Sensitivity - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
645 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
1AF00EF949D73A5F0631B2C377C03FB3

About this manual

Ion suppression is a common issue in bioanalysis by LC/MS that leads to poor recovery and increased maintenance costs. This document presents evidence that the hydroxylated surface of Agilent Bond Elut Plexa reduces interaction with endogenous materials compared to competitor products with amide residues.

The study compares the performance of Agilent Bond Elut Plexa against two competitors using human plasma spiked with beta blockers. Results show that Bond Elut Plexa delivers higher sensitivity, lower limits of detection (LOD < 0.05 ng/mL), and excellent recovery rates with good linearity.

What's inside

  • Abstract
  • Introduction
  • Materials and Methods
  • Experiment set up
  • LC Conditions
  • Results and Discussion
  • Recovery Experiment
  • Conclusion

Frequently asked questions

What is the stated limit of detection for beta blockers using Agilent Bond Elut Plexa?

The product achieves a limit of detection (LOD) between 0.01 and 0.05 ng/mL.

What is the stated limit of quantitation for beta blockers using Agilent Bond Elut Plexa?

The product achieves a limit of quantitation (LOQ) between 0.05 and 0.5 ng/mL.

What is the minimum correlation coefficient achieved for calibration curves in this study?

All compounds showed a correlation coefficient greater than 0.995.